Hioki LR5031 Mini data logger for instrumentation
The Hioki LR5031 Instrumentation Logger includes features and functions such as 7 times the recording capacity of former models, data import during recording, continuous measurement even during battery replacement, and intuitive PC software. Flexible and easy-to-use at single and multiple locations, the new LR5031 is compact and ideal for any application that requires simple set-up but long-term, reliable recording capabilities.
- For recording 4-20 mA instrumentation signals, etc.
- Stores 60,000 data points per channel
- Long-term recording capability exceeds that of previous models
- Easy-to-see dual display
- IP54 Splash-proof rating
- Data transfer during recording
- Continuous measurement even during battery replacement
- Backup function
- Intuitive PC software
- Simple set-up
- Long-term, reliable recording capabilitie
Specification Hioki LR5031 | |||
LR5000 Series Common Specifications | |||
Recording Interval | 1/ 2/ 5/ 10/ 15/ 20/ 30 seconds 1/ 2/ 5/ 10/ 15/ 20/ 30/ 60 minutes | ||
Recording Methods | One time recording - Stop recording when the memory capacity is full Endless recording - Continue recording even when the memory capacity is full (old data is overwritten) | ||
Recording Modes | Instantaneous recording - Instantaneous values are recorded at every recording interval Statistical value recording - Measure at one second intervals, and record the instantaneous, maximum, minimum, and average values within every recording interval | ||
Storage Capacity | Instantaneous value mode - 60,000 data sets per channel Statistical value mode - 15,000 data sets per channel | ||
Display Items | Measured value, Interval confi gration, Date, Time, Alarm, Remaining battery power, Number of data, Maximum data, Minimum data | ||
Recording Start / Stop | Manual or Timer | ||
Data Backup | Data from the last recording session is always backed up Back up recorded data and confi guration when battery is dead | ||
Interface | Infrared optical communications with LR5091, LR5092-20 | ||
LR5031 Specific Specifications | |||
Features | For recording 4-20 mA instrumentation signals, etc. | ||
Measurement Items | For Instrumentation / 0 to 20mA DC, 1ch | ||
Measurement Range | DC -30.00 to 30.00mA | ||
Accuracy | ±0.5%rdg. ±5dgt. (@23˚C±5˚C) | ||
Operating Range | -20˚C(-4˚F) to 70˚C(158˚F) , 80%rh or less (non-condensating) | ||
Dimensions and Mass | Approx. 79mm(3.11in)W×57mm(2.24in)H×28mm(1.10in)D 105g(3.7oz) | ||
Power Supply | LR6 (AA) Alkaline battery 1.5V×1 |