Description
Oakton, WD-35640-30, DO 450 Portable Waterproof Dissolved Oxygen Meter with Probe
DO 450 Waterproof Dissolved Oxygen Meter with Probe. Affordable optical RDO technology reduces maintenance time and costs.
Includes:
- DO 450 Meter
- Probe
Features:
- Eliminates costly, tiresome membranes and solutions
- Rugged, IP67-rated housing for plant and field applications
- USB and RS-232 output for up to 500 data sets
- 500 hours battery life, or optional universal power adapter
- Built-in stand with wall mount allows for easy benchtop measurement or storage
The RDO 450 meter uses optical measurement technology offering fast response times and long-term stability with minimal maintenance. Plus automatic temperature compensation (ATC), barometric pressure compensation, and salinity correction to further ensure measurement accuracy. The large, backlit LCD lets you view % saturation or mg/L (ppm) simultaneously with temperature.
Rugged, IP67-rated waterproof housing make the meter ideal for plant or field testing. Additional features include ready (stability) indicator, smart averaging, hold function, calibration alarm, and battery-life indicator.
Perform data logging with manual or timed collection of up to 500 data sets with real-time clock time/date stamp. Order optional data cables for USB and RS-232 connectivity.
Specification
Measures | Dissolved oxygen, °C/°F |
% saturation Range | 0.0 to 200.0% |
% saturation Resolution | 0.1% |
% saturation Accuracy | ±2.0% full scale |
mg/l (ppm) Range | 0.00 to 20.00 mg/l (ppm) |
mg/l (ppm) Resolution | 0.1 mg/l (ppm) |
mg/l (ppm) Accuracy | ±2.0% full scale |
Temperature Range | -10.0 to 110.0°C |
Temperature Resolution | 0.1°C |
Temperature Accuracy | ±0.5°C |
Temperature Calibration | Offset 0.1° increments |
ATC | Automatic or manual |
Memory | Up to 500 data sets |
Real Time Clock | Yes |
Salinity correction | Range: 0.0 to 45.0 ppt; Resolution: 0.1 ppt; Method: Key in manually and automatic correction |
Barometric pressure correction | Range: 500 to 1499 mm Hg; Resolution: 1 mm Hg; Method: Key in manually and automatic correction |
Optional AC Power | Yes |
Built-in Stand | Yes |
Dual Display | Yes |
Waterproof/Dustproof (IP67) | Yes |
Dimensions | 7 7/8" x 3 1/8" x 2 1/4" (20 x 8.3 x 5.7 cm) |